Quantity of relay outputs |
2 |
Relay commutation voltage, VDC, max. |
30 |
Relay commutation current, A, max. |
2 |
Calibration interval, months |
12 |
Ex marking, TR CU 012/2011 |
1 Ex db[ib] IIC T5 Gb X |
Ex marking, ATEX |
II 2G Ex dbib IIC T5 Gb |
Power supply range, VDC |
12…24 |
Power consumption, Wt, max. |
2 |
Ingress protection |
IP 66 |
Communication interface |
RS-485 |
Communication protocol |
Modbus |
Analog output range, mA (optional) |
4…20 |
Ambient temperature range, °С |
-40…+50 |
Weight, kg / Масса, кг |
3 |
Dimensions, mm, max. |
190 x 206 x 90 |
H2S Measurement Cell Specification (PGU-E):
Type |
Electrochemical |
Detected gas |
Hydrogen Sulphide |
Measurement range, mg/m3 (ppm) |
0…40 (0…28) |
Absolute error range, mg/m3 (within 0…10 mg/m3 measurement range) |
± 2 |
Relative accuracy, % (within 10…40 mg/m3 measurement range) |
± 20 |
Ex marking, TR CU 012/2011 |
1 Еx ib IIC Т5 Gb X |
Ingress protection |
IP 54 |
Ambient temperature range, °С |
-40…+50 |
The GSV-1E Gas Analyzer (H2S) is designed for continuous automatic measurement of the concentration of hydrogen sulphide gas in the air of the working area.
The Gas Analyzer can be supplied with a digital or analog output, or with both digital and analog outputs
Capabilities
Design features